YOUR COMPLETE SATISFACTION IS ALWAYS 100% GUARANTEED! All of our repairs are made using the highest grade, autoclaveable, OEM quality components available in the industry. Our highspeed repairs and turbine replacements are made with the absolute highest quality bearings.  Also, most of our  Highspeed repairs are completed and shipped or hand delivered within 24 - 48 hours of receipt. .

Handpiece Maintenance

· Never put handpieces with turbines in an ultrasonic cleaner.

  • · Never cool handpieces under cold running water. (Quick cool downs are bad for turbine components.)

  • · Never use a dry heat sterilizer to sterilize handpieces. (The excessive temperatures will damage bearings and other components.)

  • · Ensure that your steam sterilizer does not exceed 135C (275 F).

  • · Try not to use long and/or large head burs.

  • · For push button turbines: do not use under or oversized burs. (Undersized - .0625 diameter, Oversized - .0630 diameter. Undersized burs will slip during use and cause damage to the push button chuck. Oversized burs will stick or jam inside the spindle, causing damage to the push button spindle.)

  • · Do not put lubricant into chuck/autochuck, put cleaner only. Clean chuck weekly with proxy brush.

  • · Do not use high speed handpieces in labs, impression material debris is bad for bearings.

  • · If you use autoclave bags for handpiece sterilization, make sure you remove them from the bag when they have cooled.

  • · Check your air and water supply filters and dryers on a regular basis.

  

Lubrication Locations

Handpiece Lube locations

Handpiece Air Pressure Requirements